Transcatheter aortic valve implantation with the Evolut platform for bicuspid aortic valve stenosis: the international, multicentre, prospective BIVOLUTX registry
BACKGROUND: Prospective data about transcatheter aortic valve implantation (TAVI) in bicuspid aortic valve (BAV) patients are limited. AIMS: We aimed to evaluate the clinical impact of the Evolut PRO and R (34 mm) self-expanding prostheses in BAV patients and explore the impact of different computed tomography (CT) sizing algorithms in a prospective registry. METHODS: A total of 149 bicuspid patients were treated in 14 countries. The primary endpoint was the intended valve performance at 30 days. Secondary endpoints were 30-day and 1-year mortality, severe patient-prosthesis mismatch (PPM) and the ellipticity index at 30 days. All study endpoints were adjudicated according to Valve Academic Research Consortium 3 criteria. RESULTS: , and the mean aortic gradient was 7.2 (5.4-9.5) mmHg. No patient had more than moderate aortic regurgitation at 30 days. PPM was observed in 13/143 (9.1%) surviving patients and was severe in 2 patients (1.6%). Valve function was maintained at 1 year. The mean ellipticity index remained 1.3 (interquartile range 1.2-1.4). Overall, 30-day and 1-year clinical and echocardiography outcomes were similar between the two sizing strategies. CONCLUSIONS: BIVOLUTX demonstrated a favourable bioprosthetic valve performance and good clinical outcomes after TAVI with the Evolut platform in patients with bicuspid aortic stenosis. No impact from the sizing methodology could be identified.