Capillary versus venous haemoglobin determination in the assessment of healthy blood donors

BACKGROUND AND OBJECTIVES: To determine the accuracy of fingerstick haemoglobin assessment in blood donors, the performance of a portable haemoglobinometer (HemoCue Hb 201+) was prospectively compared with that of an automated haematology analyzer (Cell-Dyn 4000). Haemoglobin values obtained by the latter were used as the 'true' result. MATERIAL AND METHODS: Capillary fingerstick samples were assayed by HemoCue in 150 donors. Fingerstick samples from two sites, one on each hand, were obtained from a subset of 50 subjects. Concurrent venous samples were tested using both HemoCue and Cell-Dyn devices. RESULTS: Capillary haemoglobin values (HemoCue) were significantly greater than venous haemoglobin values (HemoCue), which in turn were significantly greater than venous haemoglobin values by Cell-Dyn (mean ± SD: 14.05 ± 1.51, 13.89 ± 1.31, 13.62 ± 1.23, respectively; P < 0.01 for all comparisons among groups). Nine donors (6%) passed haemoglobin screening criteria (≥ 12.5 g/dl) by capillary HemoCue, but were deferred by Cell-Dyn values (false-pass). Five donors (3%) were deferred by capillary sampling, but passed by Cell-Dyn (false-fail). Substantial variability in repeated fingerstick HemoCue results was seen (mean haemoglobin 13.72 vs. 13.70 g/dl, absolute mean difference between paired samples 0.76 g/dl). Hand dominance was not a factor. CONCLUSIONS: Capillary samples assessed via a portable device yielded higher haemoglobin values than venous samples assessed on an automated analyzer. False-pass and false-fail rates were low and acceptable in the donor screening setting, with 'true' values not differing by a clinically significant degree from threshold values used to assess acceptability for blood donation.

Capillary versus venous haemoglobin determination in the assessment of healthy blood donors | Litlas