Self-Expanding or Balloon-Expandable TAVR in Patients with a Small Aortic Annulus

BACKGROUND: Patients with severe aortic stenosis and a small aortic annulus are at risk for impaired valvular hemodynamic performance and associated adverse cardiovascular clinical outcomes after transcatheter aortic-valve replacement (TAVR). METHODS: or less in a 1:1 ratio to undergo TAVR with either a self-expanding supraannular valve or a balloon-expandable valve. The coprimary end points, each assessed through 12 months, were a composite of death, disabling stroke, or rehospitalization for heart failure (tested for noninferiority) and a composite end point measuring bioprosthetic-valve dysfunction (tested for superiority). RESULTS: ; percentage of patients with hemodynamic structural valve dysfunction, 3.5% and 32.8%; and percentage of women with bioprosthetic-valve dysfunction, 10.2% and 43.3% (all P<0.001). Moderate or severe prosthesis-patient mismatch at 30 days was found in 11.2% of the patients in the self-expanding valve group and 35.3% of those in the balloon-expandable valve group (P<0.001). Major safety end points appeared to be similar in the two groups. CONCLUSIONS: Among patients with severe aortic stenosis and a small aortic annulus who underwent TAVR, a self-expanding supraannular valve was noninferior to a balloon-expandable valve with respect to clinical outcomes and was superior with respect to bioprosthetic-valve dysfunction through 12 months. (Funded by Medtronic; SMART ClinicalTrials.gov number, NCT04722250.).